Produced by KEM, the MKC 500 series titrators provide reliable titration drawing on KEM's vast instrumentation and engineering experience.


Features:



Easy to Operate: Very simple operation just pressing [start] key

Displays measurement results with high repeatability: Guarantees below 0.3% of Relative Standard Deviation(RSD) in the measurement on 1mg H2O water-methanol standard

One-component cell can be used. (MKC-520-N and MKC-501-N) One-component cell allows for easy maintenance in replacing reagents and lower running costs

Variety of standard interface: Interface for Computer, Balance and Printer is standard equipment

Dispenser for Karl Fischer reagent. (MKC-520) The reagent dispenser as standard equipment eliminates troublesome replacements

Titration cell can be increased 2 units. (MKC-520) Once the optional stirrer and titration cell are added, two cells can be used alternatively.


Back to top of page



Pricing:




Get your order started today by calling 412-841-6988 or emailing us at



Pricing information is not currently available through pennhills-scientific.com.
Please call or email us for a detailed price quote.


Back to top of page



Technical Specifications:




Measuring Method: Karl Fischer coulometric titration method
Measuring Range: 10μg to 100mg H2O
Display resolution: 0.1μg H20
Respectability: within 0.3% RSD (n=10)/water-methanol 1mg H20
Control Method: Constant current pulse time control
Endpoint detection: AC Polarization
Display: Pre-tit(Excessive moisture state); Ready(Measurable state); Stable(Stable drift state)
Indication of endpoint: Electronic beep
Stirring method:Magnetic Stirrer; Stirring speed adjustable by 10 steps
Titration Cell: 100mL(Max. 150mL); 2 channel(option)
Drift Compensation: Automatic(Cancelable with key entry
Respectability: within 0.3% RSD (n=10)/water-methanol 1mg H20
Moisture quantity display: H2O 0.1μg to 999999μg(display at end of titration); Br2 0.9μg to 999999μg
Diagnostic Function: Electrolysis current, Measurement calue, Overtitration, Reagent life(anolyte/catholyte), Electrode contact(short/open), Parameter error, Preamplifier error, Inhibited Key entry
External Control: Printer: IDP-100 prints Date and time, Sample ID, Sample weight, H2O content, etc.; Balance: Automatic weight reading; Computer: RS-232C Interface
Ambient condition: Temperature: 5 to 35 Celsius; Humidity: below 85%RH
Power source: AC100/120/220/230/240V, 50/60Hz
Power consumption: Approximately 50W
Dimensions(mm): Main unit: 288(W) x 468(D) x 215(H); Stirrer 118(W) x 225(D) x 330(H); Solvent change unit; 240(W) x 170(D) x 405(H)
Weight: 10kg
CE Marking: EMC: EN61326


Measuring Method: Karl Fischer coulometric titration method
Measuring Range: 10μg to 100mg H2O
Display resolution: 0.1μg H20
Respectability: within 0.3% RSD (n=10)/water-methanol 1mg H20
Control Method: Constant current pulse time control
Endpoint detection: AC Polarization
Display: Pre-tit(Excessive moisture state); Ready(Measurable state); Stable(Stable drift state)
Indication of endpoint: Electronic beep
Stirring method:Magnetic Stirrer; Stirring speed adjustable by 10 steps
Titration Cell: 100mL(Max. 150mL
Drift Compensation: Automatic(Cancelable with key entry
Respectability: within 0.3% RSD (n=10)/water-methanol 1mg H20
Moisture quantity display: H2O 0.1μg to 999999μg
Diagnostic Function: Electrolysis current, Measurement calue, Overtitration, Reagent life(anolyte/catholyte), Electrode contact(short/open), Parameter error, Preamplifier error, Inhibited Key entry
External Control: Printer: IDP-100 prints Date and time, Sample ID, Sample weight, H2O content, etc.; Balance: Automatic weight reading; Computer: RS-232C Interface
Ambient condition: Temperature: 5 to 35 Celsius; Humidity: below 85%RH
Power source: AC100/120/220/230/240V, 50/60Hz
Power consumption: Approximately 50W
Dimensions(mm): 288(W) x 468(D) x 475(H)
Weight: 7.5kg
CE Marking: EMC: EN61326


Heater: Electrically conductive clear heater glass
Temperature Range: Room temp to 300 Celsius
temperature Control: Control method: proportional; Setting range: 0 to 300 Celsius; Minimum Temperature setting: 1 Celsius; Temperature sensor: Chromel-Alumel thermocouple
Display: LED Digital
Carrier gas: Nitrogen gas, governor and tubing prepared by user(Nitrogen and Air pump not included)
Gas flow: 10 to 300mL/min
Connection to KF Titrator: Can be controlled by MKC-520 with cable 980303388; key entry also available for use with other KF titrators
Power: AC100 to 240V, 50/60Hz, 150W
Dimensions(mm): 297(W) x 206(D) x 230(H)
Weight: 7kg


Heating method: Cartridge base heater
Temperature Range: Room temp to 200 Celsius
Gas flow rate: 100 to 300mL/min (Typical 200mL/min)
Carrier gas: Nitrogen gas(not included)
Power: AC100 to 120V, 50/60Hz, 400W
Dimensions(mm): 297(W) x 206(D) x 230(H)
Weight: 7kg