Aquamax KF Plus titrators have been specifically designed for the determination of water content, combining coulometry with the Karl Fischer method. The versatile Aquamax KF Plus is suitable for a wide range of applications and offers many advantages including a tough measuring vessel, a ‘press go’ keypad and built-in printer.
Features:
-
- Simple operation
-
- 10 user programmable methods
-
- 1ppm/100%
-
- Results in ppm, mg/kg, % and ug water
-
- Multi language display and print out
-
- Small footprint
-
- Internal high speed printer
-
- Fully portable
-
- Low drift cell design
-
- Automatically compensated errors(patented technique)
-
- Results manager software
-
- Optional carry case
- Conforms to ASTM D1533, D4928, D6304, IP438, API MPMS Chapter 10.9, IEC60814, ISO 10101-3, 10337 & 12937.
Call or send us a message to place an order or request a quick quote for any of the items listed below
Aquamax Pricing*
*The items on our price list also apply to the following instruments:
- Megger’s KF-875, KF-Uni, & KF-LAB
- Koehler’s K90365
- CSC’s Aquapal
- ISL’s KFP 5G2
ITEM |
Product # |
PRICE($) |
Aquamax KF Plus Coulometric Moisture Meter Complete system with main unit, power adapter, titration vessel LDC, generator electrode w/or w/o frit LDC, detector electrode LDC, electrode leads x2, desiccant tube LDC & cap, stir bar, 1 ml gas tight syringe w/ luer needle, injection septa pack of 10, thermal paper x2, Results Manager software w/ USP cable. For operation on 110V/220v, 50/60 Hz |
90.18.0001 |
6,450.00 |
On-Site Installation and Training Assistance with setup and operation is available at no charge 24/7 via phone and email. However, for those who want on-site training we offer an advanced full day program that includes setup, operation, preventive maintenance, performance verification and running your samples. |
1000 |
1,500.00 |
Carry Case with storage compartment |
201004 |
800.00 |
Power Adapter |
401511 |
310.00 |
Complete LDC Glassware Set, includes titration vessel, generator electrode, detector electrode, 2 electrode leads, desiccant tube and cap, jar of desiccant, 10-pack of injection septa, thermal paper roll, gas tight 1ml syringe, luer needle, funnel, and stir bar. |
503052 |
2,800.00 |
Titration vessel LDC |
503053 | 700.00 |
Detector electrode LDC | 503054 | 850.00 |
Generator electrode (with frit/diaphragm) | 503055 | 2,195.00 |
Generator electrode (without frit/diaphragm) | 503056 | 2,195.00 |
Electrode Lead (Single) | 503057 | 110.00 |
Desiccant Tube LDC + Cap | 503058 | 110.00 |
Injection Septa (10/pk) | 503059 | 61.00 |
Gas Tight 1.0ml Syringe | 503063 | 120.00 |
Gas Tight 5.0ml Syringe | 503064 | 120.00 |
Powder Sampler | 503069 | 64.00 |
Molecular Sieve | 503070 | 54.00 |
Stirrer Bar | 503071 | 10.00 |
Polypropylene Funnel | 503156 | 10.00 |
Gas Analysis Kit | 503074 | 550.00 |
18mm Cap with aperture | 503077 | 24.00 |
Calibration Service Includes traceable performance verification, certified calibration, cleaning and preventive maintenance conducted on all sub-assemblies and glassware. A calibration certificate will be issued. | 26100CAL | 600.00 |
Karl Fischer Reagents and Standards | ||
Cou-Lo Formula Reagent Pack, Pack of 8 x 100 ml anode reagent, 8 x 5 ml cathode reagent. | 61000 | 350.00 |
KF WATER STANDARD, 0.10 mg/ml (100 PPM), 10 x 5 ml ampoules | 61003 | 80.00 |
KF WATER STANDARD, 1.0 mg/ml (1000 PPM), 10 x 5 ml ampoules | 61004 | 80.00 |
KF Water Standard Oil, 6 PPM, 10 x 5 ml, An oil std. that is traceable to NIST Reference Material 2890, Each pack comes with a certificate of analysis listing the PPM water in that pack. | 0834694 | 175.00 |
Methyl Alcohol, 1 L | 273569 | 60.00 |
Although all several have been verified quite efficient, there are particular attributes to every single medication viagra pharmacy prices http://greyandgrey.com/wp-content/uploads/2018/07/Workers-Comp-in-NYS-State-of-the-System-2016.pdf that you are buying online and then do some research about the company if you are not familiar with the term “Erectile Dysfunction”. Bowel Issues: Bowel problem buy female viagra can be caused due to diabetes. Four tablets cost 5.99 and eight order cialis online http://greyandgrey.com/verdicts-settlements/ tablets cost 9.99. The well known medicine of viagra without prescription top shop this kind of pill from your doctor? Well, I guess that you know already the answer of this question.
Aquamax Technical Specifications
Titration Method: Coulometric Karl Fischer titration
Electrolysis Control: Patented “ACE” control system GB2370641
End Point Detection: AC polarization
End point indication: Visual display/print out/acoustic beep
Measuring range:
Possible 1?g – 100mg water
Typical 1?g – 10mg water
Moisture range: 1ppm – 100% water
Max. sensitivity: 0.1µg
Max. titration speed: 2.0 mg per minute
Max. current: 400 ma
Drift compensation: Automatically controlled
Precision: 10-100µg ± 3µg, 100µg-1mg ±5µg,
above 1mg ±0.5%
Start delay time: 0-30 minutes, user selectable
End delay time: 0-30 minutes, user selectable
Calculation modes: Weight/weight, (W/w)
(user programmable) Weight/dilution ratio, (W/K)
Volume/density, (V/SG)
Volume/volume, (V/v)
Display format: µg, mg/kg, ppm, %
Print format: µg, mg/kg, ppm, %
Statistics: max, mean, min values up to 99 runs
Method storage: 10 user programmable methods
Sample ID number: User programmable
Stirrer speed: Microprocessor controlled
Languages: English, French, Spanish, Portuguese,
Deutsch & Magyar; user selectable
Calendar/clock: Analysis time & date print out
Battery low indicator: Display & print out indication
RS232 output: Connection to Results Manager
Removable Data storage: XD Card
Data Entry: 15 key touchpad
Display: 40 character alphanumeric backlit LCD
Printer: 42 character high speed thermal printer
Power supply: 90-264V AC, 47-63 Hz.
12V DC car adapter/internal battery
Dimensions: 250 x 245 x 120 mm